Soluri A., Massari R., Trotta C., Montani L., Iurlaro G., Mangano A., Scopinaro F., Scafé R. New imaging probe with crystals integrated in the collimator's square holes . In: Nuclear Instruments & Methods in Physics Research Section A-Accelerators Spectrometers Detectors and Associated Equipment, vol. 554 pp. 331 - 9. Elsevier, 2005.
 
 
Abstract
(English)
This work presents the results obtained with a new generation portable gamma camera, assembled with an innovative technique based on CsI(Tl) scintillation crystals inserted into the square holes of a tungsten collimator. Characterization measurements have been performed on three detectors with different crystal size, in order to obtain 7×7, 8×8 and 9×9 arrays over the fixed area of 25.7×25.7 mm2, having the same overall dimensions of a Position Sensitive Photomultiplier Tube. Detector sensitivity, energy resolution and spatial resolution have been measured, and the acquired image quality has been evaluated with particular attention to pixel identification capability.
URL: http://linkinghub.elsevier.com/retrieve/pii/S0168900205017006
DOI: 10.1016/j.nima.2005.08.091
Subject Nuclear medicine imaging
Scintillation detector
Position sensitive photomultiplier tube